Main Unit Specifications |
Product | Handheld Particle Counter |
Model | Model I | Model II |
Particle Measurement | Sizes | 3 channels (0.3, 0.5, 5.0μm) | 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm) |
Flow Rate | 0.1CFM (2.83L/min) |
|
Accuracy: ±5% (Compliant with JIS B9921 and ISO21501-4) |
Maximum Measurable Concentration | 10% for 2,000,000 particles/CF |
|
(compliant with JIS B9921 and ISO21501-4) |
Counting Efficiency | 50±20% (for PSL particles near the minimum measurable size) |
|
100±10% (for PSL particles of 1.5 to 2 times as large as the minimum |
measurable size) |
|
(Compliant with JIS B9921 and ISO21501-4) |
False Count | ≦1 particle/5 minutes |
|
(Compliant with JIS B9921 and ISO21501-4) |
Size Resolution | ≦15% (for PSL particles near 0.3μm) |
|
(Compliant with JIS B9921 and ISO21501-4) |
Display | 4.3 inch color LCD, Resistive touch panel |
Communication | Standard | USB (Host: for printer and USB flash memory, Device: for PC) |
Stand (option) | RS485, Ethernet, and WiFi connection points in the base/stand |
Recording Media | Media | Internal memory |
Number and Format | Up to 10,000 records in CSV format |
Language | English, Japanese, Chinese, and Spanish |
Power | Internal Supply | Li-ion rechargeable battery (Replaceable) |
External Supply | AC adapter – Input 100 to 240 V |
Continuous Operation Time | 5 hours (Single measurement) |
Operating | Main Body | From 10 to 40 ℃, from 0 to 85%RH (With no condensation) |
Environment | Probe | From 0 to 50 ℃, from 2 to 98%RH (With no condensation) |
Dimension | W213 X H100 X D55 mm |
Weight | 650g |
Hydrothermal | Measurement Range | (Temperature) From 0 to 50℃ |
Measurement | |
(Available as an | (Humidity) From 2.0 to 98.0%RH |
option for the 6- | Display Resolution | (Temperature) 0.1℃ |
channel portable | |
particle counter) | (Humidity) 0.1% |
| Humidity Accuracy | ±3.0% (30 to 85%RH), ±5% (other range of humidity) |
| Temperature Accuracy | ±0.5℃ |
| Response Time | Approximately 60 seconds or less (90% response) |