Panther PRO Gas Leak Detector

This gas leak detector provides rapid and sensitive detection of a wide range of gases including Helium, Refrigerants and Hydrogen. With over 30 years of experience in the leak detection industry, the Panther Gas Leak Detector follows on from the success of its predecessor, the GasCheck series. Designed for the search and location of gas leaks, the Panther feels natural to hold in the hand during use. Simply draw the instrument along the area that requires testing at approximately 25mm per second. When a leak is detected, retrace the route at a slower rate until the leak is located. Once the leak is located, hold the instrument over the leak until the measurement stabilises. Internal Piezo Pump The Panther Gas Leak Detector utilises a micro thermal conductivity sensor which offers fast and sensitive detection for many gases. The internal piezo pump offers extremely stable sample flow and makes it highly stable in draughty environments. The Panther has also demonstrated its relative immunity in high electro-magnetic field environments, eg MRI scanners. ION PC & Bluetooth Connectivity Setting up the Panther Gas Leak Detector and down-loading data is simplified via the Panthers integrated Bluetooth system and ION PC Software platform. Users can select which measurement gas is required from the internal gas table, allocate the unit to an individual operator or inspection area and choose preferred operator language. Key Features : Internal Gas Table Data Logging Functionality IP44 Long Rechargeable Battery Life Bluetooth Connectivity ION PC Software Vibration, Audio & Visual Alarms Highly Resistant To Magnetic Fields
Our Products Are…
  • Economically Priced
  • Futuristic Technology
  • Guaranteed Performance with Premium Quality
  • High Reliability
SpecificationValue
DetectorMicro thermal conductivity detector (MTCD)
Battery TypeRechargeable Li-ion battery
Battery Life20 hours
Audible Alarm90 dBa at 10 cm
Factory CalibrationAt 50 % relative humidity, room temperature: 5000 ppm helium (± 5%) Leak: 0.0005 cc/s (± 5%)
Data Logging10 days continuous
Sensitivity (cc/s)Helium: 5 x 10-6
Hydrogen: 3.8 x 10-6
ResponseT90: 1 second
Flow Rate2 cc/min
Ingress ProtectionIP44
TemperatureOperating: 0°C to 50°C
Humidity0-99% R.H
Weight & Dimension447g (approx.)
320 x 80 x 55 mm (approx.)