Specifications | Details |
Carbon Dioxide Sensor Type | Solid-state Infrared |
Accuracy – Carbon Dioxide | ± 1% of reading, or ± 0.1% CO2 (whichever is greater) |
CO2 Minimum Detection Limit | 0.1% (0-50% or 0-100% range) or 0.01% (0-10% or 0-20% range) |
CO2 Calibration Controls | Digital Keypad |
CO2 Calibration Frequency | Every 12 months; calibration gas recommended. |
Expected Lifetime (oxygen) | 4-5 years |
Oxygen Range | 0.00 to 100% |
Oxygen Sensitivity | 0.01% |
O2 Resolution | 0.01% |
Analog Output | Not Available |
Resolution | 0.01% or 0.1% (user selectable) |
Accuracy | +/- 0.1% (from 0-20%), +/- 0.2% (from 20% to 50%), +/- 0.5% (from 50% to 100%) |
Drift | less than 0.4% per year |
Oxygen Sensor Type | Proprietary Electochemical |
Response Time | 20 seconds to T95 |
O2 Sensor Lifetime | 4-5 Years |
O2 Sensor Body | Anodized aluminum (CO2) and ABS Resin (O2) |
Sample Connection | Sample Probe |
Sample Method | Pump |
Sample Pump | Internal, draws ~5cc/sec, with 2-12 second adjustable timer (on side of unit) |
Pump Timing | 2-20 seconds, continuous, or cycling on/off, programmable |
Minimum Sample Required | 15cc |
Sampling Assembly | Probe with male luer connection for side-port needle, PTFE tubing |
Oxygen Calibration Controls | Digital Keypad Adjustment |
Instrument Calibration | EasyCal calibration |
Exhaust port | Vents to air |
Power Supply | 100 to 240 VAC, internationally compatibile |
Battery Operation | Included, internal shrink wrapped NiMH battery pack |
Input Voltage | 100-240V, 50-60Hz |
Battery Life | 12-16 hour runtime |
Measures: O2, CO2 | O2, CO2 |
Form Factor | Portable |
Size | 8in x 7in x 2in (203 x 51 x 178 mm) |
Weight | 2 lbs (0.9kg) |
Trace Oxygen Analyzer
This next-generation portable oxygen and carbon dioxide analyzer is designed for applications such as MAP (modified atmosphere packaging) testing, incubator monitoring, and more. The analyzer offers rapid measurements, advanced technology, and ease of use, making it a versatile and reliable choice for oxygen and carbon dioxide analysis. Key Features Advanced Technology Microprocessor-based sampling with onboard data collection. TFT full-color screen for clear and intuitive display. EasyCal calibration system for quick, touchpad-controlled calibration. Efficient Sampling and Sensor Stability Rapid results in 15-25 seconds with stable and long-life sensors. Oxygen sensor rated for 4-5 years, ensuring consistent performance. NDIR technology for CO2 sensing with no moving parts and no consumables. Ease of Use User-friendly controls allow adjustment of oxygen span, pump time, CO2 span, and more without rebooting or warmup. Durable sample probe with a liquid-blocking filter and tri-beveled, non-coring needle. Enhanced Data Management Export readings to USB in CSV format for further analysis. Portable and Durable Powered by a 12-hour internal battery for extended runtime. Includes carry case, internal pump, and accessories for convenience. Eco-Friendly and Robust Design Infrared CO2 sensor with ultra-low power consumption and virtually unlimited lifespan. Sensors unaffected by high levels of background gases like nitrogen. |