Bio Safety Cabinet

A Class II biosafety cabinet ensures a safe and controlled working environment for handling sensitive biological materials. This cabinet is designed to protect the user, the product, and the environment from potential contamination. It achieves this by maintaining a consistent inflow velocity, efficient air filtration, and strategic airflow distribution.

Key Features

Optimal Inflow Velocity: Maintains a minimum inflow velocity of 75 ft/min to ensure user and environment safety.
Efficient Air Distribution: Filtered air is evenly distributed across the work surface, ensuring consistent performance.
Dual HEPA Filtration:
70% Recirculation: Air recirculates through a supply HEPA filter to maintain cleanliness.
30% Exhaust: Air exits through an exhaust HEPA filter, ensuring minimal contamination risk.
Adjustable Air Ratio: Equipped with dampers to fine-tune the recirculation-to-exhaust ratio for optimal performance.
Enhanced Work Surface Coverage: Airflow evenly reaches areas about two to six inches above the work surface for thorough coverage.
Integrated Airflow Design: Exhaust air is drawn from the bottom of the cabinet and routed to the top, maintaining consistent airflow dynamics.

Our Products Are…
  • Economically Priced
  • Futuristic Technology
  • Guaranteed Performance with Premium Quality
  • High Reliability
Specification Details
Exterior Material Mild Steel Powder Coated
Interior Material Stainless Steel LM 304 Grades
Internal Dimensions (L × W × D) 4 × 2 × 2
Main Filter ULPA Filter (99.99% efficient at 0.3 μm particles)
Exhaust Filter HEPA Filter (99.97% efficient at 0.3 μm particles)
Pre-Filter 95% efficient at 5 microns
Clean Environment Standard Class 100
Manometer Equipped with an analog manometer
UV Light High-efficiency UV light with strong germicidal power (disinfects in 30 minutes)
Connections Gas and air connections, electrical sockets for external equipment
Front Window Tempered safety glass with UV protection
Lighting Fluorescent tube light
Door Counter-balanced sliding glass door for clear work area visibility
Mobility Fitted with castor wheels for ease of movement
Noise Reduction Equipped with specialized sound-absorbing materials
Taps Inside Unit Gas connection and 2 taps provided
Independent Switches ON/OFF switches for blower, fluorescent lamp, and UV lamp with visual status indicators
Documentation User manual, relevant test certificate for S.S. LM-304, and calibration certificate
Compliance Manufactured in compliance with GMP standards
Qualification IQ, OQ, PQ along with an informational manual